EB200 Manual
Remote control
4052.2000.02 4.14
E-7
Special characters
|
A selection of keywords with an identical effect exists for some commands.
These keywords are given in the same line and are separated by a vertical
stroke. Only one of these keywords has to be indicated in the header of
the command. The effect of the command is independent of the keywords
being indicated.
Example:
SENSe
:FREQuency
:CW|:FIXed
The two following commands of identical meaning can be formed. They set
the frequency of the device to 123 MHz:
SENSe:FREQuency:CW 123E6 = SENSe:FREQuency:FIXed 123E6
A vertical stroke in indicating the parameters marks alternative possibilities
in the sense of "or". The effect of the command is different, depending on
which parameter is entered.
Example:
Selection of parameter for command
SENSe:GCONTrol:MODE FIXed
MGC AUTO
AGC
If the parameter
FIXed
is selected, the gain is determined by
the MGC voltage. In case of
AUTO
the gain depends on the
signal.
The two parameters
MGC
and
AGC
are synonymous for
FIXed
and
AUTO
.
[ ]
Keywords in square brackets can be omitted in the header (see "Common
commands" on page 15). The device has to accept the full command
length due to reason of compatibility to SCPI standard.
Parameters in square brackets can also be optionally inserted into the
command or can be omitted.
{ }
Parameters in braces can be inserted in the command either with the
options not at all, once or several times.