DIAGnostic subsystem
4052.2000.02 4.26
E-7
. [:MEASure]:TPOint[<numeric_suffix>]? <module>
Output of the voltage at a test point on a module. With <numeric_suffix> one of the test points (1 to 8)
on the module is selected.
Parameter:
<module>
Indication of module to be queried
P1 Processor
V1 Preselection
S1 Standard
frontend
1
S2 Standard
frontend
2
Z1 IF
section
DC DC/DC
converter
SU
IF panoramic unit
Result:
Instantaneous test point voltage in mV
Example:
DIAGnostic:TPOint3? S2 -> 2163