EB200 Manual
TRIGger subsystem
4052.2000.02 4.119
E-7
4.5.19 TRIGger subsystem
The trigger subsystem is used to synchronize actions within the unit with events.
With SW option EB200CM (Coverage Measurement) triggered level measurements can be carried out.
The trigger source is selected by using command
TRIGer:SOURce EXTernal|INTernal|TIMer
.
See also Annex G (Coverage Measurement).
TRIGger[:SEQuence]
. :SLOPe
POSitive|NEGative
Setting of the slope initiating a trigger via rear-panel connector X8, pin 12 in FASTlevcw and LIST
modes
Parameters:
POSitive
trigger at positive slope
NEGative
trigger at negative slope
*RST state:
POSitive
Example:
TRIGger:SLOPe NEGative
. :SLOPe?
Query of the slope initiating a trigger via rear-panel connector X8, pin 12 in FASTlevcw and LIST modes
Parameters:
none
Result:
NEG, POS
Example:
TRIGger:SLOPe? -> NEG