MANUALE ISTRUZIONI / INSTRUCTIONS MANUAL
MTR605/025/DASM - R605/027/DASM - R605/170/DASM
Modello/Model R605DASM
E/ 1
co
mp
a
tib
ili
tà
_tubo.fm
E - C
OMPATIBILITY
W
ITH
X-
RAY
T
UBES
:
a) Compatibility between the collimator and x-ray tube is determined by the
mechanical possibility of mounting them - See the following
Table.
b) Collimator R605DASM has no inherent filtration. Check that the tube housing litera-
ture indicates congruent minimum inherent filtration (1mm or 0mm) and, that maxi-
mum radiation leakage is 30 mr/hour measured at one meter from the source when
operating at its leakage technique factors (125 at 4 mA).
c) Source values (tube housing-collimator) must not be less than 3mm Al for filtration
and must never exceed 100 mR/hr for radiation leakage. (EN 60601-1-3 par.
29.201.7/29.201.6 e 21 CFR sub-chapter J, part 1020.30 (m) (1).
d) The distance between the tube x-ray focus and the flange mounting plane (collimator
upper plate) must be 80 mm (3.14"), toll/- 1mm (0.04”).