MANUALE ISTRUZIONI / INSTRUCTIONS MANUAL
MTR605/025/DASM - R605/027/DASM - R605/170/DASM
Modello/Model R605DASM
B/ 1
d
escrizion
e_60
5DASM.fm
B - D
ESCRIPTION
Compact radiological automatic collimating system for round and elliptic fields
designed to operate with a mobile "C" arm Image Intensifier equipment.
R605DASM
The round and elliptical fields are defined as follows: The round field by 8 lead shutters
located near the exit window and a brass cone near the x-ray focus; the elliptical field
by the round field and two pairs of lead rectangular shutters located near the collimator
entrance window.
Round and elliptical field shutters are controlled by 5 stepping motors.
The collimator features a microprocessor cuit built into the collimator to control the 5
stepping-motors via external signal source with CanBus protocol.
The circuits return a CanBus protocol signal to indicate correct motor positioning.
The two pairs of lead rectangular shutters move jointly and both rotate ± 360°.