48
26.1. Local Settings
To go to local settings enter
<3.7.AnLS>
submenu.
3.7.1.Acc
Working mode
accessibility
For detailed description read section 19.2.1.
3.7.2.Aut
Averaging time
Enter this parameter to declare duration of the
process in seconds (5s, 10s, 20s, 30s, 40s, 50s,
60s) - on the basis of indications recorded within
the set time interval the scale calculates the
weighing result, i.e. an average weight value.
3.7.3.Lo
Lo threshold
For detailed description read section 19.2.4.
26.2. Animal Weighing Operation
•
Enter
<AnLS>
working mode (Animal Weighing).
•
First text
<tinnE>
is displayed for 1 s, next, window for setting duration
(in seconds) of animal weighing.
•
Press
key, parameter values are displayed successively one by one:
5[s], 10[s], 20[s], 30[s], 40[s], 50[s], 60[s].
•
Set the respective value, press
key to confirm, weighing window with
A
letter is displayed.
•
Load the weighing pan with an animal.
•
On exceeding the set mass value of
<Lo> threshold
parameter, animal
weighing starts, this is signalled with display of 'dash',
< - >
.
•
Upon process completion mass value of an animal is snapped and
displayed together with
OK
pictogram in the upper part of the display.
The snapped mass value is sent to a scale-connected printer.
•
Press key to restart animal weighing.
•
Press
key to reprint the snapped mass value.
•
Upon unloading of the weighing platform, the weighing window with letter
A
is displayed. The scale can be loaded with an animal again.
27. IMPORT / EXPORT
Function enabling you to archive weighing reports and Alibi reports, and to
copy parameters between weighing devices of the same series. Import/export
operation can be carried out by means of USB flash drive comprising
<FAT
files system>
. Upon connection of the USB flash drive to the USB A port, the
drive gets detected automatically, as a result
<IE>
submenu is created.