12.4. Automatic Internal Adjustment
Automatic internal adjustment is triggered:
•
after connection of the scale to the mains,
•
when temperature variation occurrs,
•
after passage of specified time interval.
In case of an automatic internal adjustment it is necessary to declare condition
that is to trigger the automatic adjustment. To declare the condition,
go to
<1.5.ACL>
parameter:
nonE
Adjustment disabled.
tnnP
Adjustment triggered by temperature variation greater than 3
°
C.
tinnE
Adjustment carried out in specified time intervals. For non-verified scales it
is possible to set the time interval between successive automatic
adjustment processes (read section 12.5 of this manual). For verified
scales the time interval is 12 hours.
botH
Adjustment triggered by both time and temperature.
Adjustment upon connection of the scale to the mains
•
Upon completed start-up procedure, scale stability conditions regarding
adjustment are checked, the internal adjustment is triggered automatically.
•
Adjustment process, when in progress, is signalled with display of 'dash',
< - >
.
•
Upon completed adjustment process, the scale automatically returns
to the weighing mode.
Adjustment upon temperature change
•
The scale is equipped with precise system monitoring temperature
variation and registering temperature value for each completed adjustment
process.
•
Adjustment process is triggered automatically at the moment when
temperature measured by the scale changes by more than 3°C.
•
Right before the adjustment process, scale stability conditions are
checked.
•
Adjustment process, when in progress, is signalled with display of 'dash',
< - >
.
•
Upon completed adjustment process, the scale automatically returns
to the weighing mode.