40
•
Press
key, parameter values are displayed successively one by one:
StAb
Manual printout of stable weighing result. Upon pressing
key at the
moment when the result is unstable (no
pictogram displayed), the
program first waits for the stability condition to be met, only then printout is
carried out.
nStAb
Manual printout of each weighing result. In case of unstable indication,
<?>
sign is displayed in front of the 'mass frame'.
Function available for
non-verified scales exclusively
.
rEPL
Automatic printout of the first stable weighing result above
<Lo>
threshold
(to set
<Lo>
threshold go to
<Lo>
parameter).
rEPLi
Automatic printout with time interval set in
[min]
(to set the interval go to
<Int>
parameter).
•
Press key to confirm, next go to the home screen.
19.2.3. Automatic Printout Time Interval
Parameter enabling you to set frequency of automatic printout. Printout interval
is set in minutes with 1 [min] accuracy within 1 [min] - 1440 [min] range.
Procedure:
•
Enter
<P3.Func>
menu and select given working mode.
•
Enter
<Int>
function, window for entering
time interval
value is displayed.
•
Press
key to confirm, next go to the home screen.
19.2.4. Lo Threshold
<Lo> parameter allows you to configure the function of automatic operation. In
order to save the next measurement, before carrying it out the mass indication
must get below the set net value of Lo threshold.
Procedure:
•
Enter
<P3.Func>
menu and select given working mode.
•
Enter
<Lo>
function, window for entering
Lo threshold
value is displayed.
•
Enter respective value and press
key to confirm, then continue
weighing.