11
Nominal voltage of the power supply (specified on the power supply data plate)
has to be compatible with the mains nominal voltage.
Procedure:
•
Connect the power supply to the mains. Plug it to the power supply socket
(back side of the scale housing).
•
Press
key. The key is also used to switch the scale on/off.
•
Display test proceeds (all symbols are backlit for a moment), program
name and number is displayed first, ZERO indication with reading unit
next (displayed reading unit is conditioned by scale type).
•
In case the weighing result is not zero after indication stabilisation, press
key.
6.3. Warm-Up Time
•
For correct scale operation, the workroom temperature must range
b15
°
C
÷
+30
°
C.
•
It takes 30 minutes for the device to warm up.
•
During the thermal stabilization, the indications on the screen can change.
•
If you want to carry out adjustment operation, make sure that your device
is thermally stable.
•
Temperature and humidity variations during operation may cause
indication errors. To remove the indication errors perform user adjustment.
6.4. Battery Charge Status
The scale of standard design is equipped with an internal battery. Battery state
is signalled by
pictogram, the pictogram is displayed in the top bar of the
display.
pictogram display mode
Meaning
No pictogram
Battery charged. Regular scale operation.
Pictogram displayed continuously
Too low battery charge (the scale is about to shut
down).
Charge the battery immediately.
Blinking pictogram, blink frequency:
ca. 1 s
Battery charge in progress. The device is connected
to the power supply charging the battery.
Blinking pictogram, blink frequency:
ca. 0.5 s
Battery error. Battery is damaged.