43
Communication with
the additional display
not established..
Incorrect additional display
port
set
in
<5.3.1.Prt>
parameter.
Enter < P5.ducE / 5.3.AdSP> submenu
and set correct <5.3.1.Prt> parameter
value.
Incorrect
transmission
parameters for the selected
computer port.
Enter <P4.Conn> menu and set correct
transmission parameters for the selected
additional display port.
Displayed mass unit
does not comply with
the scale data plate.
Changed scale start unit in
<9.1.UnSt> parameter.
Enter <P9.Unit / 9.1.UnSt> submenu and
set unit complying with the scale data
plate.
Changed
custom
unit
in
<9.2.Unin> parameter.
Enter <P9.Unit / 9.2.Unin> submenu and
set unit complying with the scale data
plate.
30. ERROR MESSAGES
- E r r 2 -
Value beyond zero range.
- E r r 3 -
Value beyond tare range.
- E r r 4 -
Adjustment weight or start mass out of range (
1% for adjustment weight,
10
for start mass).
- E r r 5 -
Battery error. Battery is damaged.
- E r r 8 -
Time of the following operations exceeded: taring, zeroing, start mass
determining, adjustment process.
- n u l l -
Zero value from converter.
- F U L L -
Weighing range exceeded.
- L H -
Start mass error, indication out of range (
10% of start mass).
- H i -
Display range of total mass on scale display exceeded in 'Totalizing' mode.
- E r r L o -
- Determined mass of single part in 'Parts counting' mode too small.
- Value of 'Min' threshold is greater than value of 'Max' threshold in '+/- control'
mode.
- E r r H i -
- Entered value of single part greater than maximum capacity in 'Parts
counting' working mode.
- Entered value of 'Max' threshold greater than maximum capacity in '+/-
control' mode.
- Entered reference mass greater than maximum capacity in 'Percent weighing'
mode.