17.4. Automatic Shutdown
Parameter allowing you to set time interval, in [min], after which the weighing
device shuts down automatically. If the indication is stable during the declared
time interval, the device is shut down automatically. Shutdown function is
inactive and the device cannot be turned off if any process is started or if you
operate the menu.
Procedure:
Enter
<P7.Othr / 7.4.t1>
submenu.
Press
key, parameter values are displayed successively one by one:
nonE
– function disabled,
1
,
2
,
3
,
5
,
10
.
Set the respective value, press
key to confirm, next go to the home
screen.
17.5. Date and Time
Parameter allowing you to set current date and time and to specify date and
time format.
Procedure:
Enter
<P7.Othr>
submenu and change the settings. Refer to the below
table:
Parameter
Description
<7.5.SdAt>
Enter this parameter to set current date, where the date f ormat is
YYYY.MM.DD*
.
<7.6.Stnn>
Enter this parameter to set current time, where the time format is
24H**
.
<7.7.FdAt>
Enter
this
parameter
to
select
date
format.
Available
values:
1
- DD.MM.YYYY,
2
- MM.DD.YYYY,
3
- YYYY.MM.DD* (set by default),
4
- YYYY.DD.MM.
<7.8.Ftin>
Enter this parameter to select time format. Available values:
24H**
(set by
default),
12H**
.
*)
- Date format: Y
– year, M – month, D – day.
**)
- Time format: 12H
– 12-hour format, 24H - 24-hour format.
17.6. Default Operator Settings
Parameter allowing you to restore default operator settings.