17
Adjustment has to be carried out:
prior weighing,
if long breaks between successive measuring series occur,
if the ambient temperature has changed dynamically,
if the scale has been relocated.
Adjustment types:
external adjustment,
<1.1.CA-E>
, performed using an external weight
of declared mass, i.e. mass that cannot be modified,
user adjustment,
<1.2.CA-u>
, performed using an external weight of mass
of any value comprised within the weighing range, however not lower than
30% of the maximum capacity value.
12.1. External Adjustment
External adjustment must be carried out using an external adjustment weight
of class F
1
.
Procedure:
Enter
<P1.CAL / 1.1.CA-E>
submenu, text
<UnLoAd>
(remove weight)
is displayed.
Remove the load from the weighing pan and press
key.
Mass of an empty weighing pan is determined, this is signalled with
display of 'dash',
< - >
. Next, text
<LoAd>
(load weight) and mass
value that is to be loaded, e.g.
2000g
(conditioned by scale type), are
displayed.
Load the weighing pan with weight of specified mass value and press
key.
Weight mass is determined, this is signalled with display of 'dash',
< - >
. Next, text
<UnLoad>
(remove weight) is displayed.
Remove the load form the weighing pan,
<1.1.CA-E>
submenu
is displayed.
12.2. User Adjustment
User adjustment must be carried out using an external adjustment weight
of class F
1
.