37
23.2. Reference Sample Mass Determined by Weighing
Enter
<3.4.dEu / 3.4.2.UUt>
submenu, set
<
S_S
>
value.
Enter
<dEu>
working mode (Percent weighing), first, text
<LoAd>
is displayed for 1 second, then the weighing window.
Load the weighing pan with the reference sample. When the indication
is stable (
pictogram is displayed) press
key to confirm the mass.
Mass of the weighed load is automatically set as reference sample mass,
the home screen is displayed along with
100.000%
value.
23.3. Reference Sample Mass Determined by Entering the Mass Value
Enter
<3.4.dEu / 3.4.2.UUt>
submenu, set
<Suu>
value.
Enter
<dEu>
working mode (Percent weighing).
First text
<SEt_Ut>
is displayed for 1 s, next, window for declaring mass
value of the reference sample.
Enter the value and press
key to confirm. The home screen
is displayed automatically with
0.000%
value.
If the value of entered reference sample mass is greater than
max capacity value, then message <Err Hi> is displayed.
24. WORKING MODE - PEAK HOLD
Peak Hold is a working mode allowing you to snap value of maximum force
applied to the weighing pan during one weighing process.
24.1. Local Settings
To go to local settings enter
<3.5.toP>
submenu.
3.5.1.Acc
Working mode accessibility
For detailed description read section 19.2.1.
3.5.2.Lo
Lo threshold
For detailed description read section 19.2.4.
24.2. Peak Hold Operation
Enter
<3.5.toP / 3.5.2.Lo>
submenu, set
<Lo>
parameter value
(Lo threshold) after exceeding of which maximum force is to be registered.