- 79 -
16.
COOPERATION WITH A PRINTER OR A COMPUTER
CAUTION
A peripheral device that is connected to RS 3232 output of a balance, has
to be powered from the common low voltage power network equipped with
common anti-shock protection in a way to preclude possible occurrence of
different potentials in zero cables of the peripheral device and the balance.
Transmission parameters programmed in a balance:
-
Baud rate
- 2400 – 19200 bit / s
-
Data bits
- 8
-
Stop bit
- 1
-
Parity control
- none.
The value indicated on balance display can be sent through RS 232 output
to a peripheral device in one of four accessible ways:
•
manually
– on pressing
key
•
automatically
– on stabilization of weighing result
•
continuously
– on activation of a function or sending a
command
•
on command sent from a peripheral device (see additional functions).
Value indicated on balance’s display can be sent through RS232 output in
the following form:
•
stable
– data is sent immediately on stabilization of weighing
result
•
unstable
– on pressing of
key causes immediate sending
of display status to a peripheral device (on a printout
such status is marked with
<?>
symbol
proceeding the
weighing result).