- 28 -
10.3.
Autozero function
In order to ensure balance’s precise mass indication, autozero (
Auto
) software
parameter has been introduced. The application of this function is automatic
control and correction of zero indication. When function is enabled, it
compares balance indications at declared time interval e.g. 1 s, on conditions
that weighing pan is unloaded and display indication is close to zero.
If results vary less than declared AUTOZERO range e.g. one division, balance
will zero automatically, display marker of stable measurement result
and
precise zero marker
.
If AUTOZERO function is enabled, then each weighing process starts from
precise zero point. There are, however, some cases when this function can be
a disturbing factor of measuring process; for instance very slow placing of load
on the weighing pan (e.g. load pouring) – in such case system of zero
indication correction can also correct actual indication of loaded mass.
OFF
- autozero disabled
On
- autozero enabled.
Go back to weighing mode with procedure of saving changes
(see point - 8. 2. – return to weighing).
Содержание APP 10/2C
Страница 2: ...2 NOVEMBER 2012...