background image

Ref No.

052017 

2. Power on: Press “ON/MEM” button. The thermometer is ready for use after you see the 

  icon on the screen and hear two beep sounds.

 

   

In this mode, you can get the body temperature by forehead measurement.

 

3. Temperature taking:   
    3.1 Body temperature: 

Remarks: 1. Forehead temperature is displayed in oral mode. This mode converts the forehead temperature to display its   

“oral equivalent.”     

                  2. Before the measurement, please stay in a stable environment for 5mins and avoid exercise, bath for 30mins. 

                                  3. Remember to keep the forehead area clean and away from sweat, cosmetics and scar while taking temperature. 

4. The "Clinical Bias” is -1.4 ~ -1.7°C. 

5. The "Limits of Agreement” is 0.98. 
6. The "Repeatability” is 0.20°C. 

 

3.1.1 Press the ON/MEM button to power on the device. 

Forehead mode is the default mode.

 You can see the 

  icon on the 

screen and hear two beep sounds (See Figure 1). In this mode, you can hold the thermometer within 1 cm from the central 
forehead (See Figure 2) and press the "Start" button to get the forehead measurement. The time consuming for measurement 
might be 1 second. After two short beeps are heard, the 

  icon stops flashing to be ready for next measurement. 

3.2 Get the room temperature: 

3.2.1 For room temperature, the thermometer should be placed on the table and avoid the direct sunshine or air conditioner flow on.   
3.2.2 When in power off, room temperature will always be shown on the screen with    icon. 
3.2.3 The room temperature can be reference after 15min later. 

4. After Measurement: 

4.1 Power off: Device will automatically shut off if left idle for more than 1 minute to extend battery life. 

The LCD will only show the date, time and room temperature.

 

4.2 Clean the probe after each use to ensure an accurate reading and avoid cross contamination.    (See the section of Care and Cleaning for details.) 

 

 

Important Notes 

Care and cleaning

 

The probe is the most delicate part of the thermometer. Use with care when cleaning the probe lens to avoid damage.   

a. After each use, use the 70% alcohol swab or the soft cotton moistened with the 70% alcohol to clean the Probe and Lens. 
b. Allow the lens to fully dry for at least 1 minute. 

c. Keep the unit dry and away from any liquids and direct sunlight.   
d. Storage temperature range: It should be stored at room temperature between -20~+50°C, RH

85% 

  Holding the thermometer too long may cause a higher ambient temperature reading of the probe.   

This could make the body temperature measurement lower than usual.

 

Battery replacement 
 

 

 

Disposal of the used batteries

Please comply with the local regulation. 

This device is supplied with one lithium cell CR2032x1.

 

1

  Open the battery cover: Insert a pointed object, such as a pen, into battery cover pick hole. At the same time, use thumb to push battery 

cover out.   

2

  Hold the device and flip the battery out with a small screwdriver   

3

  Insert the new battery under the metal hook on the left side and press the right side of the battery down until you hear a click   
* The positive (+) side up and the negative (-) side pointed down.   

4

  Replace the battery cover 

 

 

 

Trouble shooting: 

Error Message 

Problem 

Solution 

 

Error 5~9, the system is not functioning properly. 

Unload the battery, wait for 1 minute and repower it. If the message 

reappears, contact the retailer for service. 

 

Measurement before device stabilization.

 

Wait until all the icons stops flashing.

 

 

The ambient temperature is not within the range between 10

o

C and 40

o

C (50

o

F ~104

o

F). 

Allow the thermometer to rest in a room for at least 15 minutes at room 
temperature: 10

o

C and 40

o

C (50

o

F ~104

o

F). 

 

Temperature taken is higher than +42.2

o

C (108°F). 

Please select the target within specifications. If a malfunction still exists, 

please contact the nearest retailer. 

 

Temperature taken is lower than +34

o

C (93.2°F). 

 

Device cannot be powered on to the ready stage. 

Change with a new battery. 

 

Warranty 

Warranty: 12 months   

Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)   

Ex.SN:E512A000001, the first “E” is External, the second number “5” is the last number of manufacture year, the third and the fourth number “12” is the manufacture month, the others is the serial number. 
 

Note: The thermometer is calibrated at the time of manufacture. If at any time you question the accuracy of temperature measurements, please contact the dealers or nearest 

service address. 

 

Warning: No modification of this equipment is allowed. 

 

Please read the instructions for use       

  BF type applied part 

 

Radiant Innovation Inc.    Http://www.radiantek.com.tw 
Add: 1F, No.3, Industrial East 9

th

 Road, Science-Based Industrial Park, 

HsinChu, Taiwan 300. 

Medical Technology Promedt Consulting GmbH   
Add: Altenhofstrasse 80, D-66386 St. Ingbert, 

Germany 

Figure 2 

Figure 2 

Отзывы: