a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM 

and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to 
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the 
TH5xy series  is used exceeds the applicable RF compliance level above, the  TH5xy series  should be observed to verify normal operation. If 
abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the  TH5xy series. 

b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than  3  V/m. 

 

Guidance and manufacturer’s declaration – electromagnetic immunity 

The  TH5xy series  is intended for use in the electromagnetic environment specified below. The customer or the user of the  TH5xy series  should 
assure that it is used in such an environment. 

Immunity test

 

IEC 60601 test level

 

Compliance level

 

Electromagnetic environment 

– 

guidance

 

Electrostatic                   
discharge (ESD) 
IEC 61000-4-2 

6 kV contact 
8 kV air 

6 kV contact 
8 kV air 

Floors should be wood, concrete or ceramic tile. If floors are 
covered with synthetic material, the relative humidity should be at 
least 30 %. 

Electrical fast 
transient/burst 
 
IEC 61000-4-4 

2 kV for power supply lines 
1 kV for input/output lines 

Not applicable 

Mains power quality should be that of a typical commercial 
or hospital environment. 

Surge 
IEC 61000-4-5 

1 kV line(s) to line(s) 
2 kV line(s) to earth 

Not applicable 

Mains power quality should be that of a typical commercial 
or hospital environment. 

interruptions and 
voltage variations 
on power supply 
input lines 
 
IEC 61000-4-11 

<5 % UT 
(>95 % dip in UT) 
for 0,5 cycle 

 

40 % UT 
(60 % dip in UT) 
for 5 cycles 

 

70 % UT 
(30 % dip in UT) 
for 25 cycles 

 

<5 % UT 
(>95 % dip in UT) 
for 5 sec 

Not applicable 

Mains power quality should be that of a typical commercial 
or hospital environment. If the user of the  TH5xy series 
requires continued operation during power mains 
interruptions, it is recommended that the  TH5xy series  be 
powered from an uninterruptible power supply or a battery. 

Power frequency 
(50/60 Hz) 
magnetic field 
 
IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic fields should be at levels 
characteristic of a typical location in a typical commercial or 
hospital environment. 

NOTE UT is the a.c. mains voltage prior to application of the test level. 

 

Recommended separation distances between 

portable and mobile RF communications equipment and the  ME EQUIPMENT or ME SYSTEM

 

The  TH5xy series  is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the 
user of the  TH5xy series  can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF 
communications equipment (transmitters) and the  TH5xy series  as recommended below, according to the maximum output power of the 
communications equipment.

 

Rated maximum output power 

of transmitter 

Separation distance according to frequency of transmitter 

m

 

150 kHz to 80 MHz 

d = 

1,2 

 

80 MHz to 800 MHz 

d = 

1,2 

 

800 MHz to 2,5 GHz 

d = 

2,3 

 

0,01

 

N/A 

0,12 

0,23 

0,1

 

N/A

 

0,38

 

0,73

 

1

 

N/A

 

1,2

 

2,3

 

10

 

N/A

 

3,8

 

7,3

 

100

 

N/A

 

12

 

23

 

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

in meters (m) can be estimated 

using the equation applicable to the frequency of the transmitter, where 

is the maximum output power rating of the transmitter in watts (W) 

according to the transmitter manufacturer.   
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, 
objects and people.

 

 