4.1
Indications of use
The Heart Guardian HR-503 is indicated for use on victims of cardiac arrest who
are exhibiting the following signs:
- Unconscious
- Not breathing
- Without circulation
The Heart Guardian HR-503 is intended for use by personnel who have been
trained in its operation. Users should have received training in basic life support /
AED, advanced life support or a physician-authorized emergency medical
response training program.
The Heart Guardian HR-503 is intended for use with same electric pads on both
adult and child with a separate button operation (adult mode on patients greater
than 8 years old or over 25 kg, child mode on patients between 1 and 8 years of
age or up to 25 kg). Default is set as adult mode when AED is released.
It is highly recommended for every potential user to complete periodic training
for CPR and HR-503 operation. Please contact national/local representative for
Heart Guardian series to find out detailed CPR and AED operating training
schedule
Heart Guardian, HR-503 is an emergency medical device used for patient who
stops breathing and lapsed into unconsciousness due to sudden cardiac arrest.
Sudden cardiac arrest patient shows symptoms as below;
No response
to the external stimulus
Unusual or no breath
Do not use AED if patient responds or conscious