Warning and Caution
Please refer to the following for safety use of HR-503
Certain contents may be indicated repeatedly in other parts of manual
Do not use if the patient shows any reaction or is conscious
HR-503 is a portable defibrillator manufactured for ventricular
fibrillation patients. Prior to AED operation, it is very important
to check patient’s consciousness and respiration condition; do not
use if the patient is conscious and reacting
HR-503 is set as “Adult Mode” as default
In adult mode, AED delivers about 150J of electric shock to
patient. In child mode, about 50J of electric shock is set to be
delivered. If the patient is under the age of 7 or weighting below
25kg, please set “Child Mode” by using “Adult/child select
button”.
Be careful of electric shock hazard
Electric shock can be delivered to operator or bystander
through unexpected ways. To avoid this hazard during AED
operation, keep followings:
Do not use AED in standing water, underwater, rain,
snowing. Move patient to a dry area
Confirm and instruct anyone to stay away from the
patient prior to pressing ‘Shock Button’
When patient is contacted with any conductive materials such
as metals, stay away from the conductive materials
Do not use around flammable materials
To avoid possible fire or explosion hazard, do not use around the
following situation/place.
close to volatile gas or oil
close to hyperbaric oxygen tank (within 10m)
close to hydrogen tank (within 10m)
close to flammable materials
Do not attempt to recharge or cause heat/damage the battery
(BT-503)
Recharging or heat/damage on battery(BT-503) may
cause an explosion or fire hazard. If the battery is low, please
exchange to a new one