Place where is exposed to direct sunlight
Wet or flooded places
Near a heater or heavy electrical devices
Places where excessive shock or shake may occur
Near flammable materials
Places exposed to the extreme environmental weather condition
5.1.6
Battery and AED pads maintenance
①
Battery maintenance
If the battery is low, AED may not be operated properly in emergency situation.
When status indicator shows X, battery pack shall be replaced immediately.
Periodically
check manufacturing date indicated on battery label, and replace
to new one if it is expired (4 years from manufacturing date)
Replacing battery pack
Pull out the battery pack to be replaced on backside of the AED,
by pressing a lever on left side of the battery pack
Push in the new battery pack facing the label down to hear ‘click’
sound. When battery is installed, AED starts a self-diagnosis and
turns into a standby mode