User Manual 3162A/B
Maintenance and Performance Checks 5-25
6
RD – U37 pin 16
Negative going 400 ns width pulses
7
WR – U37 pin 17
Negative going 400 ns width pulses
8
LDACK – U37 pin 5
Negative going 6.5
µ
s width pulses
9
ASIC LCS – U91 pin 26
Negative going 6.5
µ
s width pulses
10
LR/W - U91 pin 18
Negative going 60
µ
s width pulses
Output Amplifier and
Amplitude Control
Checks
Problems with the output amplifier and amplitude control circuits can be
detected if the Model 3162B accepts word serial commands and responds
correctly to word serial queries, but fails to output the correct waveform with
correct amplitude and offset levels. If there is a problem with the output
amplifier or with the amplitude control circuit, proceed with the checkout
procedure given in Table 5-2.
Before starting the following procedure, reset the Model 3162B to the factory
defaults using the *RST command. Then send the command:
: :OUTP ON.
:INST:SEL B
:OUTP ON
Table 5-2, Output Amplifier and Amplitude Control Checks
S t e p
Engine Board A Test Point
Expected Result
1
P1 connector (output amplifier input)
600 mVp-p sinewave, 1 MHz
2
U5, pin 6
Approximately 1 Vp-p sinewave, +6 V offset
3
U6, pin 6
Approximately 1 Vp-p sinewave, -6 V offset
4
U4, pin 2
-7 V dc
5
U2, pin 2
+7 V dc
Engine Board B Test Point
6
U13, pin 6
1.2 Vp-p sinewave, 1 MHz
7
U4, pin 6
Approximately 1 Vp-p sinewave, +6 V offset
8
U5, pin 6
Approximately 1 Vp-p sinewave, -6 V offset
9
U3, pin 2
-7 V dc
10
U1, pin 2
+7 V dc
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com