User Manual 3162A/B
Maintenance and Performance Checks 5-10
Sine Characteristics
Specified Total Harmonic Distortion:
<0.5% from 10.00 Hz to
100.0 kHz, (measured with minimum 4096 horizontal points and sampling
rate of 500 MS/s).
Specified Harmonic Signals:
-50dBc to 100 kHz
-40dBc to 10 MHz
-25dBc to 200 MHz
Equipment:
Distortion Analyzer, Spectrum Analyzer, 50
Ω
feedthrough
termination and 20dB attenuator.
1.
Configure the Model 3162B as follows:
Function:
User Waveform
Frequency:
100 kHz
Amplitude:
3 V
Remote Commands:
:RES
:FUNC:MODE USER
:TRAC:DEF 1,5000
:TRAC:SEL 1
:FREQ:RAST 500e6
:VOLT 3
:OUTP ON
:INST:SEL B
:FUNC:MODE USER
:TRAC:DEF 1,5000
:TRAC:SEL 1
:VOLT 3
:OUTP ON