3156B User Manual
Racal Instrument © 2005
Maintenance and Performance Checks 6-25
Output:
On
Test Procedure
1. Verify burst FM operation on the oscilloscope as follows:
Waveform:
Burst of 5 Sine cycles
Sine
Frequency:
10kHz
Burst
Period: 1ms
Max
A:
1.5MHz
Min
A:
500kHz
Test Results
Pass
Fail
FM – Std
Waveforms, Gated
Run Mode
Equipment
: Oscilloscope (LeCroy LT342, fitted with jitter package),
function generator
Preparation
:
1. Configure the oscilloscope as follows:
Time Base:
0.1 ms
Sampling Rate: 50MS/s at least.
Trace A View: Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude:
1V/div
2. Connect 3156B Channel 1 output to the oscilloscope input,
channel 1
3. Connect the 3156B SYNC output to the oscilloscope input,
channel 2
4. Configure the function generator as follows:
Frequency 1kHz
Run
Mode: Continuous
Waveform:
Squarewave.
Amplitude:
Adjust to TTL level on 50
Ω
5. Connect the function generator output connector to the 3156B
TRIG IN connector
6. Configure the 3156B channels 1/2 as follows:
Reset
Wave Mode: Modulation
Modulation Mode: FM
Modulation Run Mode: Gate
Carrier Frequency: 1MHz
Modulating
Frequency:
10kHz
Deviation:
500kHz
Enable
Source:
Hardware
Sync:
On
Output:
On
Содержание 3156B
Страница 5: ......
Страница 6: ...This page was left intentionally blank...
Страница 50: ...3156B User Manual Configuring The Instrument 2 6 Racal Instrument 2005 This page was left intentionally blank...
Страница 119: ...3156B User Manual Racal Instrument 2005 WaveCAD 4 27 Figure 4 11 The FSK Data Table...
Страница 370: ...3156B User Manual Maintenance and Performance Checks 6 42 Racal Instrument 2005 This page was left intentionally blank...
Страница 374: ...3156B User Manual Product Support 7 4 Racal Instrument 2005 This page was left intentionally blank...