3156B User Manual
Racal Instrument © 2005
Maintenance and Performance Checks 6-19
2. Remove the cable from 3156B channel 1 and connect to channel
2
3. Repeat the test procedure as above for channel 2
Test Results
Pass
Fail
NOTE
Leave the same setup for the next test
Step Advance
Equipment
: Oscilloscope, function generator
Preparation
:
1. Use the same preparations as for step advance, except change
mode to single sequence advance
2. Change Oscilloscope configuration to single:
Test Procedure
1. Press the manual trigger button on the function generator and
observe that one cycle waveform advances through the
sequence table repeatedly with each external trigger signal. Note
that you need to press the Single mode on the oscilloscope for
each trigger advance
Test Results
Pass
Fail
2. Remove the cable from 3156B channel 1 and connect to channel
2
3. Repeat the test procedure as above for channel 2
Test Results
Pass
Fail