1269 User Manual
Using The Enhanced Monitoring System 4-24
Selecting The Trigger
Delay Application
The trigger delay may be placed between ONE of the following:
1)
TRIG IN input to selected TTLTRG line
2)
Selected TTLTRG line to TRIG OUT output
3)
TRIG IN input to TRIG OUT output
4)
Nothing (no delay)
The following commands may be used to select the delay
application:
VXI:CONF:MONITOR:TRIG:DELAY:STATE INPUT
VXI:CONF:MONITOR:TRIG:DELAY:STATE OUTPUT
VXI:CONF:MONITOR:TRIG:DELAY:STATE NONE
The first command applies the delay between the TRIG IN
input and the selected TTLTRG line. The second command
places the delay between the selected TTLTRG line (or TRIG
IN input) and the TRIG OUT output. The third command
removes any trigger delay from the trigger subsystem.
The present trigger delay application may be read using the
query:
VXI:CONF:MONITOR:TRIG:DELAY:STATE
The reply to this query will be one of the following:
INP
OUTP
NONE
Programming The
Front Panel Display
The EMS has a 10 character LED display which is capable of
displaying all printable characters. This includes both
uppercase and lowercase “A” through “Z”, the comma, single
quote, double quote, forward and backward slash, asterisk,
dollar sign, percent sign, and so on. A space character may
also be displayed.
At Power-On and after a reset command (“*RST”), the front
panel display is turned on. The front panel will display the
message:
Standby
if the AC line power switch is off. If the AC line power switch is