6 Maintenance and cleaning
6.1 Maintenance
The RIDA
®
ABSORBANCE 96 is maintenance-free. Each time the instrument is switched on,
an internal self-test is carried out to ensure, that there are no malfunctions.
To check the linearity and accuracy of the measuring instrument for routine quality controls, a
reference plate, e.g. from Hellma GmbH (item code 666-R113) can be used.
There are no parts within the RIDA
®
ABSORBANCE 96 that can be serviced by the customer.
It is only necessary to ensure that the device is kept clean.
6.2 Cleaning
Remove the Micro-USB cable from the RIDA
®
ABSORBANCE 96 before cleaning.
Warning!
Make sure before cleaning that there is no microtiter plate in the
RIDA
®
ABSORBANCE 96.
Cleaning the housing
The surfaces of the housing should be cleaned regularly. For this purpose, you can use e.g. a
cloth or a sponge lightly wetted in water. For heavier soiling, clean the surface of the housing
with a mild soap solution diluted with water or glass cleaner and then wipe with a lightly
moistened cloth or sponge to remove any residue.
Do not use scouring agents.
Cleaning the slot
To clean the slot of the RIDA
®
ABSORBANCE 96, first open the four screws on the bottom of
the instrument (see Figure 5) using the hex key delivered separately.
Содержание RIDA ABSORBANCE 96
Страница 1: ...R Biopharm AG...
Страница 2: ...RIDA ABSORBANCE 96 Version 1 0 Bedienungsanleitung User manual 2...
Страница 4: ...RIDA ABSORBANCE 96 Version 1 0 Bedienungsanleitung User manual 4 Bedienungsanleitung...
Страница 31: ...RIDA ABSORBANCE 96 Version 1 0 Bedienungsanleitung User manual 31...
Страница 32: ...RIDA ABSORBANCE 96 Version 1 0 Bedienungsanleitung User manual 32 User manual...