![R-Biopharm RIDA ABSORBANCE 96 Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/r-biopharm/rida-absorbance-96/rida-absorbance-96_user-manual_3074780054.webp)
RIDA
®
ABSORBANCE 96 Version 1.0 | Bedienungsanleitung/User manual
54
7 Performance data and technical data
Tab. 5:
Performance data and technical data
Parameter
Value
Type of product
Absorbance reader
Product name
RIDA
®
ABSORBANCE 96
Service life
10 years at an average use of 4 h/day
Housing material
Aluminum
Place for use
Laboratory
Degree of contamination
2
Temperature (storage/measurement) 5 - 45 °C
Temperature (transport)
-10 - 50 °C
Relative humidity
(storage/measurement/transport)
0 - 90 %
Air pressure (measurement)
900 - 1070 hPa
Air pressure (storage/transport)
600 - 1070 hPa
Height
Max. 5100 m
Measurement method
Absorbance
Light source
LED
Measuring technique
Endpoint and kinetic
Measurement channels
96
Filter
4
Wavelength
405, 450, 540, 630 nm
Measuring range
0 - 3.5 OD
Resolution
0.001 OD
Accuracy*
0.000 - 2.
000 OD ≤ (1 % + 0
.010 OD)
2.000 - 3.
000 OD ≤ (1
.5 % + 0.010 OD)
Reproducibility**
0.000 - 2.
000 OD ≤ (0
.5 % + 0.005 OD)
2.000 - 3.
000 OD ≤ (1
% + 0.010 OD)
Linearity***
0.000 - 2.
000 OD ≤ 1 %
2.000 - 3.
000 OD ≤ 1
.5 %
Connection to computer
USB 2/USB 3 with 5 VDC and max. 3 A
Dimensions
55 x 96 x 154 mm
Содержание RIDA ABSORBANCE 96
Страница 1: ...R Biopharm AG...
Страница 2: ...RIDA ABSORBANCE 96 Version 1 0 Bedienungsanleitung User manual 2...
Страница 4: ...RIDA ABSORBANCE 96 Version 1 0 Bedienungsanleitung User manual 4 Bedienungsanleitung...
Страница 31: ...RIDA ABSORBANCE 96 Version 1 0 Bedienungsanleitung User manual 31...
Страница 32: ...RIDA ABSORBANCE 96 Version 1 0 Bedienungsanleitung User manual 32 User manual...