background image

19 

[F1 PT -] 

Stop periodic probe temperature reporting. 

[F1 CT -] 

Stop periodic sample holder temperature reporting. 

[*E+] 

Enable appearance of a warning dialog if a main window 

button  
 

 

or control is used while a script is running. 

[F1 PX -] 

Reset display of probe temperature to 0.1 °C precision. 

[*MSG + The multiramp script run is complete] 

 

 

Notify user, with beeping. 

 

The script sets a target temperature of 10 degrees and then turns temperature control on and waits until the 
user is satisfied with the state of equilibration.  During this time the probe temperature is reported every 30 
seconds (the sample holder temperature is reported every 30 seconds throughout the run).  The script then 
initiates the first ramp from 10 to 40 degrees at 4 degrees/min with the computer beeping each time a probe 
temperature is reported at 2 degree intervals.  The second ramp from 40 to 45 degrees is slower (0.2 
degrees/min), with the computer beeping every 0.5 degrees (to provide more information over a critical range 
of temperatures).  The third ramp takes the temperature quickly to 80 degrees, again with beeping at 2 degree 
intervals.  At this point beeping is stopped and an 8 minute delay of script execution allows time for 
equilibration at 80 degrees; the probe temperature is reported every 30 seconds.  The final ramp is down to 20 
degrees at 2.5 degrees/min with probe temperature reports every 5 degrees.  This is followed by a 7 minute 
delay for equilibration with automatic probe temperature reports every 30 seconds.  Figure 1 shows a typical 
example of the sample holder and probe temperatures as a function of time. 

 

 

Actual data obtained by running the example script in the QNW Serial Control Program 
(V2.3).  The sample holder temperatures are plotted as a solid line without tokens; the 
probe temperatures are shown as filled circles connected by a line. 

 

 

Содержание qpod 2e

Страница 1: ...way Avenue Suite 4 Liberty Lake WA 99019 8606 Tel 509 624 9290 Fax 509 624 9488 E mail quantum qnw com Web www qnw com Instruction Manual qpod 2e Temperature Controlled Sample Compartment for Fiber Op...

Страница 2: ...luded with Your Purchase 2 System Setup 3 Software Installation 4 Optics 4 System Operation 6 Error Conditions 7 Temperature Specifications 7 Other Specifications 7 Optical slits provided with the qpo...

Страница 3: ...ter provides temperature control and variable speed magnetic stirring and holds optical slits for limiting light access to the cuvette as needed A dry gas purge may be used to minimize condensation wh...

Страница 4: ...d gas on the side of the qpod 2e A flow of gas must be used to prevent condensation on the faces of the cuvette when working below the dew point temperature Set the dry gas flow rate to 50 200 cc min...

Страница 5: ...ample to roughly double the amount of emitted light collected An imaging lens assembly contains two inch diameter 30 mm focal length broadband AR coated fused silica lenses The principal point of one...

Страница 6: ...ding the black end piece screw the silver sleeve back into place 9 Insert the SMA connector and silver sleeve into the hole in the optical assembly For a first adjustment insert the assembly until the...

Страница 7: ...ng the focus of the light This adjustment is most important at UV wavelengths Measurements can be made without these alignments However in the unaligned system the reference spectrum may change with t...

Страница 8: ...rror message may not be returned immediately after the command containing the error was sent the text of the bad command is included bracketed by and Error 09 replies will be apparent only when the Sc...

Страница 9: ...s 8 5 mm from the bottom surface of the cuvette Thus for some kinds of cuvettes it is necessary to specify this standard z height Lenses Collimating optics use a single 6 mm diameter 12 mm focal lengt...

Страница 10: ...1 Sample holder is a four position turret with probe capability Assigned Identities ID 10 single cuvette holder 11 single cuvette holder with probe capability 12 high temperature single cuvette holder...

Страница 11: ...the target temperature F1 TT Turn off automatic reporting of manual changes to the target temperature Instrument Status F1 IS What is the current instrument status F1 IS 0 S Response is four parameter...

Страница 12: ...c probe temperature report F1 PA S 0 5 Automatically report probe temperature every increment of 0 5 degrees during a ramp Increment must be a positive value without sign in tenths between 0 1 and 9 9...

Страница 13: ...Note For the higher ramp settings the observed rate may be lower than that calculated from RT and RS or it may be nonlinear over part of the temperature range because the maximum possible rate of heat...

Страница 14: ...to send a long series of commands by loading a text file known as a controller script that can be easily generated in a text editor such as Windows Notepad 9 Provides a number of program specific scri...

Страница 15: ...efined in the first line of the script see the example script in Appendix 3 where the second line sets INTERVAL to 0 6 seconds or 0 01 minutes WD This program command will cause the script to Wait for...

Страница 16: ...of a script LCT or LCT Enable or Disable listing of Current Temperature replies during running of a script LPT or LPT Enable or Disable listing of Probe Temperature replies during running of a script...

Страница 17: ...ata collected to that time CTD This command will clear the time temperature records In addition the time will be restarted at zero and a new record of time temperature will be started MSG message or M...

Страница 18: ...rature periodically LTT Do not list target temperature returns in script window LCT Do not list current temperature returns in script window LPT Do not list probe temperature returns in script window...

Страница 19: ...e reporting is already running BPT Turn off the computer beep each time a probe temperature report is received D 800 Wait 8 min to allow temperatures to stabilize Ramp back to 20 F1 RT S 25 Set Rampin...

Страница 20: ...e intervals The second ramp from 40 to 45 degrees is slower 0 2 degrees min with the computer beeping every 0 5 degrees to provide more information over a critical range of temperatures The third ramp...

Страница 21: ...re monitoring and parameter acquisition be done by the same program Lacking that then one would like to be able to acquire temperature and parameter data at the same time interval with both acquisitio...

Страница 22: ...he temperature as a function of time Panel B shows the average absorbance as a function of time Panel C combines the data in A and B to show the change in the average absorbance as a function of tempe...

Отзывы: