background image

 

SYSTEM OPERATION 

1.

 

Use a liquid sample of at least 1.5 ml and place 
it in a standard 1 x 1-cm square cuvette.  Place 
the cuvette in the cuvette holder in the center 
of the 

qpod 2e

.  Standard microcuvettes may 

also be used for smaller volumes.  The 

qpod 2e

 

is designed to hold microcuvettes in which the 

“z” dimension (distance between the bottom 

of the cuvette and the optical center line) is 8.5 
mm. 

 

Note: 

The 

qpod 2e

 is intended to hold a cuvette 

with a standard wall thickness of 1.25 mm and total 

width of 12.5 mm.  Cuvettes with unusually thick 

walls will not properly fit and may damage the 

holder. 

 

2.

 

Place the magnetic stir bar in the cuvette.  
Stirring speed will be set by the program Q-
Blue. 

3.

 

If you wish to monitor the temperature inside 
the cuvette, plug a standard Series 400 or 
Series 500 thermistor probe into the ¼ - inch 
phone jack located on the corner of the qpod 
2e to the left of the front panel.  Thread the 
probe through the split rubber plug in the qpod 
2e cap and place it in the cuvette out of the 
way of the light. 

4.

 

Provide a source of light and a detector. 

5.

 

Use the 5/64-inch hex driver from the 
accessory kit to maximize the signal by 
adjustments of the three fiber steering screws 
(

Figure 3

).  This adjustment will make small 

movements of the position of the fiber end 
relative to the lens.  Also, loosen the set screw 
to insert the silver sleeve deeper into the lens 
assembly.  Inserting the silver sleeve further 
into lens assembly moves the end of the fiber 
closer to the lens, altering the focus of the 
light.  This adjustment is most important at UV 
wavelengths. (Measurements can be made 

without these alignments.  However, in the 
unaligned system the reference spectrum may 
change with temperature and must be re-
measured each time the set temperature is 
changed.  Careful alignment will minimize such 
artifacts.) 

6.

 

Use the optical slits or blanks (provided in the

 

qpod 2e

 accessory kit) around the cuvette in a 

manner that correctly limits the excitation and 
emission light. 

7.

 

Put the cover and cap on the 

qpod 2e

8.

 

Turn on the submersible pump by plugging it 
in.  Cooling water flows through a heat 
exchanger and removes heat from the 
thermoelectric device when the temperature 
of the holder is being lowered.  Insufficient 
flow will allow the heat exchanger temperature 
to rise.  If the temperature of the heat 
exchanger exceeds a certain cutoff value, 
temperature control will be automatically shut 
down to protect the holder. 

Cooler circulating water can result in improved 
performance of the cuvette holder at low 
temperatures, whereas warmer water can 
improve performance at very high 
temperatures.  

Do not use warm or room 

temperature water when setting 
temperatures near room temperature, as 
temperature instabilities are likely to arise.

  

Cool the circulating water by the addition of ice 
to the bucket. 

A refrigerated circulating bath may be used to 
provide circulating fluid to the 

qpod 2e

.  Using 

such a bath, temperatures well below the 
specified -30 °C lowest temperature may be 
readily achieved.  Contact us for suggestions on 
how to achieve very low temperatures. 

9.

 

Plug power from the AC adapter into the qpod 
2e front panel. 

10.

 

Connect the qpod 2e to your computer using 
the USB cable. 

Содержание qpod 2e

Страница 1: ...way Avenue Suite 4 Liberty Lake WA 99019 8606 Tel 509 624 9290 Fax 509 624 9488 E mail quantum qnw com Web www qnw com Instruction Manual qpod 2e Temperature Controlled Sample Compartment for Fiber Op...

Страница 2: ...luded with Your Purchase 2 System Setup 3 Software Installation 4 Optics 4 System Operation 6 Error Conditions 7 Temperature Specifications 7 Other Specifications 7 Optical slits provided with the qpo...

Страница 3: ...ter provides temperature control and variable speed magnetic stirring and holds optical slits for limiting light access to the cuvette as needed A dry gas purge may be used to minimize condensation wh...

Страница 4: ...d gas on the side of the qpod 2e A flow of gas must be used to prevent condensation on the faces of the cuvette when working below the dew point temperature Set the dry gas flow rate to 50 200 cc min...

Страница 5: ...ample to roughly double the amount of emitted light collected An imaging lens assembly contains two inch diameter 30 mm focal length broadband AR coated fused silica lenses The principal point of one...

Страница 6: ...ding the black end piece screw the silver sleeve back into place 9 Insert the SMA connector and silver sleeve into the hole in the optical assembly For a first adjustment insert the assembly until the...

Страница 7: ...ng the focus of the light This adjustment is most important at UV wavelengths Measurements can be made without these alignments However in the unaligned system the reference spectrum may change with t...

Страница 8: ...rror message may not be returned immediately after the command containing the error was sent the text of the bad command is included bracketed by and Error 09 replies will be apparent only when the Sc...

Страница 9: ...s 8 5 mm from the bottom surface of the cuvette Thus for some kinds of cuvettes it is necessary to specify this standard z height Lenses Collimating optics use a single 6 mm diameter 12 mm focal lengt...

Страница 10: ...1 Sample holder is a four position turret with probe capability Assigned Identities ID 10 single cuvette holder 11 single cuvette holder with probe capability 12 high temperature single cuvette holder...

Страница 11: ...the target temperature F1 TT Turn off automatic reporting of manual changes to the target temperature Instrument Status F1 IS What is the current instrument status F1 IS 0 S Response is four parameter...

Страница 12: ...c probe temperature report F1 PA S 0 5 Automatically report probe temperature every increment of 0 5 degrees during a ramp Increment must be a positive value without sign in tenths between 0 1 and 9 9...

Страница 13: ...Note For the higher ramp settings the observed rate may be lower than that calculated from RT and RS or it may be nonlinear over part of the temperature range because the maximum possible rate of heat...

Страница 14: ...to send a long series of commands by loading a text file known as a controller script that can be easily generated in a text editor such as Windows Notepad 9 Provides a number of program specific scri...

Страница 15: ...efined in the first line of the script see the example script in Appendix 3 where the second line sets INTERVAL to 0 6 seconds or 0 01 minutes WD This program command will cause the script to Wait for...

Страница 16: ...of a script LCT or LCT Enable or Disable listing of Current Temperature replies during running of a script LPT or LPT Enable or Disable listing of Probe Temperature replies during running of a script...

Страница 17: ...ata collected to that time CTD This command will clear the time temperature records In addition the time will be restarted at zero and a new record of time temperature will be started MSG message or M...

Страница 18: ...rature periodically LTT Do not list target temperature returns in script window LCT Do not list current temperature returns in script window LPT Do not list probe temperature returns in script window...

Страница 19: ...e reporting is already running BPT Turn off the computer beep each time a probe temperature report is received D 800 Wait 8 min to allow temperatures to stabilize Ramp back to 20 F1 RT S 25 Set Rampin...

Страница 20: ...e intervals The second ramp from 40 to 45 degrees is slower 0 2 degrees min with the computer beeping every 0 5 degrees to provide more information over a critical range of temperatures The third ramp...

Страница 21: ...re monitoring and parameter acquisition be done by the same program Lacking that then one would like to be able to acquire temperature and parameter data at the same time interval with both acquisitio...

Страница 22: ...he temperature as a function of time Panel B shows the average absorbance as a function of time Panel C combines the data in A and B to show the change in the average absorbance as a function of tempe...

Отзывы: