background image

 

6.

 

Q-Blue 

CD

 with 

temperature control software, 

help file and this qpod 2e manual.

 

C.  BATH 100 

1.

 

submersible pump 

2.

 

plastic bucket 

3.

 

a length of tubing 

D.  Optional Items

 

- If you ordered

 any optical 

components separately, such as polarizers or 
optical filter holders, you will find them packed in a 
separate internal box. 

SYSTEM SETUP 

1.

 

If desired, fasten the qpod 2e down to a table 
or optical breadboard.  The holes in the 
corners of the base are located and sized such 
that it will attach via ¼-20 screws to a table 
with 1-inch centers or with M-6 screws to a 
table with 25 mm centers. 

2.

 

Supply circulating water to the Peltier unit.  
Attach tubing from the submersible pump to 
either one of the water hose barbs on the qpod 
2e.  Attach another piece of tubing from the 
qpod 2e back to the bucket.  Put water in the 
bucket, turn on the pump by plugging it into an 
outlet and check to be sure that water is 
flowing back into the bucket.  Check for leaks.  
Unplug the pump until you have completed the 
remainder of the system set up. 

3.

 

If you plan to work at low temperatures, 
connect a source of dry gas (typically nitrogen) 
using a leng

th of tubing with 1/8” (3mm) inside 

diameter, to the hose barb labeled “gas” on 

the side of the qpod 2e.  A flow of gas must be 
used to prevent condensation on the faces of 
the cuvette when working below the dew point 
temperature.  Set the dry gas flow rate to 50 - 
200 cc/min. 

SOFTWARE INSTALLATION 

The 

qpod 2e

 may be controlled using a USB cable.  

If you will be using USB to control the 

qpod 2e

 you 

must install the software before connecting the 
hardware to the computer, otherwise the 
sequence is not important.   

4.

 

Insert the Q-Blue software CD into the drive.  If 
the installation does not start automatically, 
locate the Setup.exe file in the root directory 
and run it.  The installation process starts with 
a small black window that is shown during 
installation of the drivers needed to control the 

qpod 2e

 through a USB connection.  This 

window will then be replaced by the software 
installation window.  Follow the onscreen 
prompts to complete the installation. 

Setting up for temperature control using USB

 

5.

 

Connect the 

qpod 2e

 to your computer using 

the USB cable provided.  The New Hardware 
installation process will begin automatically 
and take a few moments to finish. 

OPTICS 

There are two basic optical configurations for the 

qpod 2e

: one to measure 

absorbance

 and the 

other to measure 

fluorescence

.   

Optics to measure absorption  

 

 

Figure 1

 - Absorbance configuration 

As shown in 

Figure 1

, to set the 

qpod 2e

 up for 

absorbance measurements insert collimating lens 

Содержание qpod 2e

Страница 1: ...way Avenue Suite 4 Liberty Lake WA 99019 8606 Tel 509 624 9290 Fax 509 624 9488 E mail quantum qnw com Web www qnw com Instruction Manual qpod 2e Temperature Controlled Sample Compartment for Fiber Op...

Страница 2: ...luded with Your Purchase 2 System Setup 3 Software Installation 4 Optics 4 System Operation 6 Error Conditions 7 Temperature Specifications 7 Other Specifications 7 Optical slits provided with the qpo...

Страница 3: ...ter provides temperature control and variable speed magnetic stirring and holds optical slits for limiting light access to the cuvette as needed A dry gas purge may be used to minimize condensation wh...

Страница 4: ...d gas on the side of the qpod 2e A flow of gas must be used to prevent condensation on the faces of the cuvette when working below the dew point temperature Set the dry gas flow rate to 50 200 cc min...

Страница 5: ...ample to roughly double the amount of emitted light collected An imaging lens assembly contains two inch diameter 30 mm focal length broadband AR coated fused silica lenses The principal point of one...

Страница 6: ...ding the black end piece screw the silver sleeve back into place 9 Insert the SMA connector and silver sleeve into the hole in the optical assembly For a first adjustment insert the assembly until the...

Страница 7: ...ng the focus of the light This adjustment is most important at UV wavelengths Measurements can be made without these alignments However in the unaligned system the reference spectrum may change with t...

Страница 8: ...rror message may not be returned immediately after the command containing the error was sent the text of the bad command is included bracketed by and Error 09 replies will be apparent only when the Sc...

Страница 9: ...s 8 5 mm from the bottom surface of the cuvette Thus for some kinds of cuvettes it is necessary to specify this standard z height Lenses Collimating optics use a single 6 mm diameter 12 mm focal lengt...

Страница 10: ...1 Sample holder is a four position turret with probe capability Assigned Identities ID 10 single cuvette holder 11 single cuvette holder with probe capability 12 high temperature single cuvette holder...

Страница 11: ...the target temperature F1 TT Turn off automatic reporting of manual changes to the target temperature Instrument Status F1 IS What is the current instrument status F1 IS 0 S Response is four parameter...

Страница 12: ...c probe temperature report F1 PA S 0 5 Automatically report probe temperature every increment of 0 5 degrees during a ramp Increment must be a positive value without sign in tenths between 0 1 and 9 9...

Страница 13: ...Note For the higher ramp settings the observed rate may be lower than that calculated from RT and RS or it may be nonlinear over part of the temperature range because the maximum possible rate of heat...

Страница 14: ...to send a long series of commands by loading a text file known as a controller script that can be easily generated in a text editor such as Windows Notepad 9 Provides a number of program specific scri...

Страница 15: ...efined in the first line of the script see the example script in Appendix 3 where the second line sets INTERVAL to 0 6 seconds or 0 01 minutes WD This program command will cause the script to Wait for...

Страница 16: ...of a script LCT or LCT Enable or Disable listing of Current Temperature replies during running of a script LPT or LPT Enable or Disable listing of Probe Temperature replies during running of a script...

Страница 17: ...ata collected to that time CTD This command will clear the time temperature records In addition the time will be restarted at zero and a new record of time temperature will be started MSG message or M...

Страница 18: ...rature periodically LTT Do not list target temperature returns in script window LCT Do not list current temperature returns in script window LPT Do not list probe temperature returns in script window...

Страница 19: ...e reporting is already running BPT Turn off the computer beep each time a probe temperature report is received D 800 Wait 8 min to allow temperatures to stabilize Ramp back to 20 F1 RT S 25 Set Rampin...

Страница 20: ...e intervals The second ramp from 40 to 45 degrees is slower 0 2 degrees min with the computer beeping every 0 5 degrees to provide more information over a critical range of temperatures The third ramp...

Страница 21: ...re monitoring and parameter acquisition be done by the same program Lacking that then one would like to be able to acquire temperature and parameter data at the same time interval with both acquisitio...

Страница 22: ...he temperature as a function of time Panel B shows the average absorbance as a function of time Panel C combines the data in A and B to show the change in the average absorbance as a function of tempe...

Отзывы: