Page 42
9530/8530 Manual Version 4.5
:INFOrmation
?
Query only. Returns model,
serial number, firmware
version, and FPGA version
numbers. The same as the
*IDN? Command.
:NSID
?
Query only. Returns firmware
and FPGA identification
numbers.
:CAPS
0/1 or OFF/ON
The default value is 1, which
means the unit is not case
sensitive. 0 means the
commands sent to the unit
must be capitalized.
*Note: To change this
parameter the unit must be
power cycled before the
command will take effect.
IEEE
488.2 Common Commands
Command:
Parameter Range:
Notes:
*IDN
?
Query only. Returns model, serial number, firmware version, and FPGA
version numbers.
*RCL
1 - 24
RECALL
*RST
RESET. Resets parameters only same as *RCL 0
*SAV
1 - 24
SAVE
*TRG
Creates a soft trigger input.
*TTG
Creates a soft trigger for the trigger input only. Used when the dual trigger
option is enabled.
*GTG
Creates a soft trigger for the gate input only. This command is only active
when the dual trigger option is enabled.
*LBL
?
Used to query the label of the last saved or recalled configuration.
String Value
Command to attach a string label to the current settings. The string must
be in double quotes and no longer than 14 characters. Command must be
followed by a *sav [1/2/n] command to take effect.
*Note: To see the label on the screen a display update or reboot must take
place.
*ARM
0/1 or OFF/ON
Resets all channel counters simultaneously when the channels are in
either single shot or burst mode.
*Note: The system must be in continuous mode (this command is
functionally the same as pressing the Run/Stop button).