Page 40
9530/8530 Manual Version 4.5
Keyword:
Command:
Sub-
Command:
Parameter Range:
Notes:
:PULSe[1/2/n]
Command to change the units’ channel
specific settings.
:STATe
0/1 or OFF/ON
Enables/Disables output pulse for selected
channel.
:WIDTh
10[ns] -
999.99999999975[s]
Sets the pulse width for the selected channel.
The command should be sent without units. If
for example 50ns is desired the parameter
sent should be 50e-9, or the decimal
equivalent.
:DELay
-99.99999999975[s]
to
999.99999999975[s]
Sets the delay from the timing reference to
when the pulse is created. The command
should be sent without units. If for example
50ns is desired the parameter sent should be
50e-9, or the decimal equivalent.
:SYNC
T0,CHA,CHB-CHH
Allows the user to select the timing reference
for each channel.
*Note: When in external clock input mode T
0
will be the clock input.
:MUX
0-255
Decimal representation of an 8 bit binary
number (example: 255 = 1111 1111)
:POLarity
NORMal /
COMPlement /
INVerted
Normal is active HIGH, Inverted and
Complement are active LOW.
:OUTPut
Command to change the channels’ output
parameters.
:MODe
TTL / ADJustable
Allows the user to select either TTL logic
mode or Adjustable voltage output mode.
:AMPLitude
2.0 to 20[V]
Allows the user to select the voltage potential
for Adjustable output mode.
:CMODe
NORMal / SINGle /
BURSt / DCYCle
Allows the user to select the pattern of outputs
to use on the channel level.
:BCOunter
1 to 9,999,999
When the channel is in Burst mode will allow
user to select the number of pulses to output
with each input clock pulse.
*Note: The commas should be omitted.
:PCOunter
1 to 9,999,999
When the channel is in duty cycle mode will
allow the user to select the number of pulses
to create with each input clock pulse.
*Note: The commas should be omitted.
:OCOunter
1 to 9,999,999
When the channel is in duty cycle mode will
allow the user to select the number of pulses
to suppress with each input clock pulse.
*Note: The commas should be omitted.
:WCOunter
0 to 9,999,999
Allows the user to select how many clock
cycles to wait until the channel should start
creating an output pulse.
*Note: The commas should be omitted.
:CTRIg
TRIG / GATE
Sets the Trigger source for the channel
selected.
*Note: For the gate to be used as a trigger
source the unit must have the dual trigger
option.