Bit Error Rate Tester | BERT 1005
Quantifi Photonics Ltd.
Version
2.04
61
Syntax
:SLOT<n>:GROUP:ED?
Description
Query the ED channels within the specified group
Parameters
No parameters
Response
A value or string.
Example
:SLOT5:GROUP:ED? -> 1
Command
:SLOT<n>:GRouP:TEMPerature?
Syntax
:SLOT<n>:GRouP:TEMPerature? ALL
Description
Query the temperature of all groups of channels
Parameters
MIN:
Returns the minimum temperature of the module
MAX:
Returns the maximum temperature of the module
ACT:
Returns the actual temperature of the module
ALL:
Returns all the above values in a comma separated string
Response
Depending on the parameters, the response will be a string, or a comma separated string
of values
Example
:SLOT5:GROUP:TEMPERATURE? ALL -> 46.97
9.5.2
Configuration commands
Command
:CALCulate<n>:DATA<m>:EALarm?
Syntax
:CALCulate<n>:DATA<m>:EALarm?<wsp><STATE|DATA|LOCK|ERROR|COUNT|BITS|B
ER|FULL|INFO>
Description
Query the ED error measurement.
Parameters
STATE: Query the state of the ED measurement.
DATA: Query the presence of the data pattern.
LOCK: Query the frequency lock status.
COUNt|ERROR: Query the total error count.
BITS: Query the total number of bits.
BER: Query the bit error rate.
FULL: Query the information specified by all the parameters.
INFO: Provides a description of what the return values represent.
By default, the command Query the total error count.
Response
The requested value, or a comma separated array. When
FULL
is requested, the
format is "
<STATE>,<DATA>,<LOCK>,<COUNt|ERROR>,<BITS>,<BER>
".
The
STATE
parameter returns an integer response corresponding to the following:
-1: The ED has not been run by the user.
0: The ED has been stopped by the user.
1: The ED is currently running.
2: The ED has reached the error overflow.
3: The ED has been stopped automatically, data is no longer recognizable by the ED.
4: The ED cannot start as the data is not recognized by the ED.
Example
:CALCulate3:DATA2:EALarm? FULL 1->
1,1,1,5.11,7.03,6.07
:CALCulate3:DATA2:EALARM? INFO ->
-1:INVALID
0:STOPPED (USER)
1:RUNNING
2:OVERFLOW
3:STOPPED (AUTO)
4:INVALID SIGNAL
Command
:CALCulate<n>:DATA<m>:EALarm:ELAPsed?
Syntax
:CALCulate<n>:DATA<m>:EALarm:ELAPsed?
Description
Query the running time for the ED measurement.
Parameters
No parameters
Response
The number of seconds the ED measurement has been running.
Example
:CALCulate3:DATA2:EALarm:ELAPsed? -> 7917.7923
Command
:CALCulate<n>:DATA<m>:EDIagram?
Содержание BERT 1005 Series
Страница 1: ......
Страница 6: ...Bit Error Rate Tester BERT 1005 Quantifi Photonics Ltd Version 2 04 6 Service and repairs 93 ...
Страница 94: ...2 04 ...