Bit Error Rate Tester | BERT 1005
Quantifi Photonics Ltd.
Version
2.04
47
8.3.7
Specify the result display type
The BER measurement can be displayed either by the accumulated BER information or by the
instantaneous BER information.
•
ACCUMULATED BER
–
Indicates that the BER value is calculated from the beginning of the
measurement. The BER value is equal to the ratio between the error counter and the bit counter
since the beginning of the measurement.
•
INSTANTANEOUS BER
–
Indicates that the BER value is calculated from the last point of the
data reading. The BER value is equal to the ratio between the error counter and the bit counter
since the last point of reading. The counters are reset after each reading.
To set the result display type, click the RESULT DISPLAY drop-down list and select the display mode.
8.3.8
ED point mode
The ED point mode of the error detector can be enabled by changing the mode to MANUAL using the
drop-down list. When operating in manual ED point mode, the voltage and phase decision levels for the
error detector can be modified.
When operating in manual mode, the maximum ERROR count is 2
16
-1.