Bit Error Rate Tester | BERT 1005
Quantifi Photonics Ltd.
Version
2.04
65
Description
Set the synthesizer frequency to an arbitrary value within the valid operating range.
Parameters
value: The desired new value
MIN: Set to the minimum value
MAX: Set to the maximum value
DEF: Restores the default value
Response
No response
Example
:OUTP3:CLOC:FREQ:ARB 7.96
Command
:OUTPut<n>:CLOCk:FREQuency:STanDard?
Syntax
:OUTPut<n>:CLOCk:FREQuency:STanDard?[<wsp><MIN|MAX|DEF|SET|LIST|ALL|
STEP|UNIT>]
Description
Query the current synthesizer frequency standard settings.
Parameters
MIN: Query the minimum configurable value
MAX: Query the maximum configurable value
DEF: Query the default power-on value
SET: Query the current SET value
LIST: Query the list of discrete options for configurable values (if available)
ALL: Query all the values specified by MIN, MAX, DEF, SET, and LIST
STEP: Query the minimum resolution of the frequency value
UNIT: Query the default unit for the frequency value
By default, this command Query the current SET value.
Response
A single value or a comma separated string of values. For ALL, the format of the
response is "<MIN>,<MAX>,<DEF>,<SET>,<LIST>".
If the frequency is set to an arbitrary value, the SET value in the response is NAN.
Example
STANDARD MODE:
:OUTP3:CLOC:FREQ:STD? -> 14.5
ARBITRARY MODE:
:OUTP3:CLOC:FREQ:STD? -> NAN
Command
:OUTPut<n>:CLOCk:FREQuency:STanDard
Syntax
:OUTPut<n>:CLOCk:FREQuency:STanDard<wsp><value|MIN|MAX|DEF>
Description
Set the current synthesizer frequency to a standard value.
Parameters
value: The desired new value
MIN: Set to the minimum value
MAX: Set to the maximum value
DEF: Restores the default value
Response
No response
Example
:OUTP3:CLOC:FREQ:STD 14.5
Command
:OUTPut<n>:CLOCk:POWer?
Syntax
:OUTPut<n>:CLOCk:POWer?[<wsp><MIN|MAX|DEF|SET|LIST|ALL|STEP|UNIT>]
Description
Query the power of the RF output clock
Parameters
MIN: Query the minimum configurable value
MAX: Query the maximum configurable value
DEF: Query the default power-on value
SET: Query the current SET value
LIST: Query the list of discrete options for configurable values (if available)
ALL: Query all the values specified by MIN, MAX, DEF, SET, and LIST
STEP: Query the minimum resolution of the power value
UNIT: Query the default unit for the power value
By default, the command Query the current SET value.
Response
A single value or a comma separated string of values. For ALL, the format of the
response is "<MIN>,<MAX>,<DEF>,<SET>,<LIST>".
Example
:OUTPut3:CLOCk:POWer? -> 0
Command
:OUTPut<n>:CLOCk:POWer
Syntax
:OUTPut<n>:CLOCk:POWer<wsp><value|MIN|MAX|DEF>