Operating Procedures
5-16
PyroMark Q96 ID User Manual 01/2016
5.5.2
Loading the reagent cartridge
1.
Open the dispensing unit cover: Release the latch and
then open the cover.
2.
Insert the filled reagent cartridge into position. It will only
fit in one orientation; the label should be facing the user.
3.
Close the dispensing unit cover. Ensure the latch snaps
into its locked position.
4.
Close the instrument lid.