Maintenance
PyroMark Q96 ID User Manual 01/2016
7-11
Procedure:
1.
Ensure that no vacuum is applied to the Vacuum Prep
Tool, i.e. the vacuum switch is closed (Off), and the
vacuum pump is switched off.
2.
Disconnect the vacuum pump from the mains power.
3.
Remove the broken tubing at one end and empty any
liquid waste into an empty beaker.
4.
Disconnect the other end of the tubing and discard the
tubing and any liquid waste.
5.
Cut the new vacuum tubing into three pieces and
assemble it. Use the old tubing as a template for the size
of the new tubing pieces. Ensure that the tubing is
connected to the pump’s vacuum fitting.
6.
Reconnect the vacuum pump to the mains power.
Содержание PyroMark Q96 ID
Страница 1: ...Version 5 January 2016 Sample Assay Technologies PyroMark Q96 ID User Manual...
Страница 2: ......
Страница 8: ...Contents Contents 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 14: ...Safety Information 1 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 32: ...Installation Procedures 4 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 58: ...Operating Procedures 5 26 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 78: ...Maintenance 7 12 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 96: ...Appendix A A 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 106: ...Appendix C C 2 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 114: ...Appendix D D 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 122: ...Appendix E E 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 125: ......
Страница 126: ...1100446 01 2016 Sample Assay Technologies...