7.7 REPLACING PCB LITHIUM BATTE Y
R
te personnel.
T
u
O
g
b
•
U
•
Lift off the lid.
located at the top right hand edge of the main
circuit board.
See Fig-23
.
•
o
ose of the battery according to local
re
•
R
It is recommended that the replacement of the battery is only
carried by Purite approved Engineers or by Puri
The Battery has a 5-year life, but it is recommended that it
should be changed every 3 years.
•
urn off the water supply,
isolate from the electrical supply and
nplug the mains lead from the rear of the unit.
•
pen the front door to gain access to the two Top cover retainin
utton head socket screws.
nscrew using a 3mmAF Allen key.
•
The Lithium battery is
To remove the battery gently pries up the spring retainer and slide
ut he battery.
See Fig-24
. Disp
gulations.
efit the new battery ensuring the +ve side is facing upwards
Incorrect fitting of the battery could cause irreversible
damage to the main PCB. ve side of battery is
facing upwards. Always use recommended battery Part No.
R083085
.
•
R
er back on and turn on
the water supply.
Fig-23
Fig-24
eplace top cover, switch the electrical pow
Main
PCB
Battery
Spring
Retainer
Clip
Select Neptune Manual Iss: B
Содержание SELECT NEPTUNE Analytical
Страница 45: ...Select Neptune Manual Iss B...
Страница 46: ...Select Neptune Manual Iss B...
Страница 72: ...7 9 PREVENTATIVE MAINTENANCE AND CONSUMABLE REPLACEMENT GUIDE Select Neptune Manual Iss B...
Страница 82: ...Select Neptune Manual Iss B...
Страница 87: ...Select Neptune Manual Iss B...
Страница 88: ...Select Neptune Manual Iss B...
Страница 89: ...Select Neptune Manual Iss B...
Страница 90: ...Select Neptune Manual Iss B...