The cartridge can now be disposed of, refer to
Section 9.4
for
•
To refit a new cartridge simply remove it from its packaging.
Be
o black plugs used to seal
o
).
•
Then slide the pack back into the holder. The locking lever will
“Click” indicating the pack is secure.
•
If a new RHS pack has been replac
confirm that a new pack has been fitted followed by a request to
reset the next exchange date, just follow the simple on screen
commands.
•
NB
If a new pack has not located correctly, the sensors located
in the pack holder will set off an alarm and raise a corresponding
alarm message on the screen, eg.,
“No LHS Pack”
. Refer to
Section 8 Troubleshooting
should this occur.
•
Turn on the water supply.
.2
REPLACING THE INTERNAL 0.2µm FILTER (Analytical)
•
Firstly if the unit is still in “
Processing
” mode, select
STOP
, to
conds.
automatic air bleed connection.
only.
With the air bleed connector detached unscrew both of the water
retaining clip and discard.
DO NOT over tighten the water connections as this may damage
NB: Check that the filter has been refitted with the arrow on the
•
al supply and turn on the water.
rompts which will guide you through the process.
advice on disposal.
sure to
remove the tw
ports
,
(see Fig-8
the tw
ed the unit should ask you to
7
place the unit into Power On mode.
•
Allow the unit to depressurise for approximately 30 se
•
Switch off the unit and isolate the mains water supply.
•
To gain access to the filter the right hand side cover will need to
be removed.
To remove the cover, unscrew the two retaining screws located
•
at the top and bottom front edge of the side cover.
•
Once released the cover can be pulled away from the unit.
•
The filter is located at the front and has water connections top
and bottom and one connection on the side. This is the
•
Gently unscrew, anti-clockwise, the air bleed connection; holding
the clear fitting
•
connections.
•
With all of the connectors detached pull the filter from its
•
Un-pack the new filter and reconnect the air bleed line and water
lines.
•
the fittings.
•
side of the filter pointing upwards.
Switch on the electric
•
You will now need to reset the filter expiry date , to do this follow
on screen p
Select Neptune Manual Iss: B
Содержание SELECT NEPTUNE Analytical
Страница 45: ...Select Neptune Manual Iss B...
Страница 46: ...Select Neptune Manual Iss B...
Страница 72: ...7 9 PREVENTATIVE MAINTENANCE AND CONSUMABLE REPLACEMENT GUIDE Select Neptune Manual Iss B...
Страница 82: ...Select Neptune Manual Iss B...
Страница 87: ...Select Neptune Manual Iss B...
Страница 88: ...Select Neptune Manual Iss B...
Страница 89: ...Select Neptune Manual Iss B...
Страница 90: ...Select Neptune Manual Iss B...