Step
First stage of
clean: Recirculation of
lean g chemicals for 30
5:
in
minute
c
Clean Recirc
Time Left 34:59
Process Mimic
Note
: mimic changes to
indicate presence of “
CP”
packs in place of NCP
s
.
Step 6:
of recir
starts t
minutes.
packs.
Use
V
U
to bleed filter.
After 30 minutes
culation the unit
o rinse for 5
Clean Rinse
Time Left 05:00
Process Mimic
Step 7:
rinsing
depressur
safe remo
cleaning packs.
Note: En
storage ta
sure if fed from
nk that there is
approx 15 litres of water
present to rinse the unit.
After 5 minutes of
the unit
ises to allow
val of the
CLEANING
Depressure
Step 8:
Remove and
discard used cleaning
packs and refit either th
original NCP packs or fi
e
t
new ones.
CLEANING
Remov
ks
e cleaning Pac
Fit Both NCP Packs
New RHS Pack Fitted?
NO YES
If “
YES
” you will be
instructed to reset the
pack expiry date. See
Section 7.1
Step 9:
Once all the
consumables have been
fitted press Exit to return to
Proce
ssing mode.
CLEA
ED
N COMPLET
Press EXIT
EXIT
Select Neptune Manual Iss: B
Содержание SELECT NEPTUNE Analytical
Страница 45: ...Select Neptune Manual Iss B...
Страница 46: ...Select Neptune Manual Iss B...
Страница 72: ...7 9 PREVENTATIVE MAINTENANCE AND CONSUMABLE REPLACEMENT GUIDE Select Neptune Manual Iss B...
Страница 82: ...Select Neptune Manual Iss B...
Страница 87: ...Select Neptune Manual Iss B...
Страница 88: ...Select Neptune Manual Iss B...
Страница 89: ...Select Neptune Manual Iss B...
Страница 90: ...Select Neptune Manual Iss B...