PureAire Monitoring Systems, Inc.
5
11.
Audio Horn
— This optional built-in horn is a 90dB high pitched audio
sound that will activate when the oxygen levels go below the selected alarm
thresholds. The audio alarm is non-latching and will automatically turn off
when the oxygen levels go above the alarm thresholds
NOTE
:
The audio alarm is an immediate alarm. Oxygen levels must recover
above the alarm thresholds before the horn turns off. There is no
alarm delay function available.
1.2.5 Transmitter Interior
1.
Power Analog Terminal Block
— This terminal block is where the 24VDC
power and 4-20 mA analog output connection is made.
NOTE:
PureAire has added additional contacts for +24VDC power and Common to
accommodate additional wiring for remote horns and strobes
2.
Sensor Cell Connector —
This connector is where the Oxygen sensor cell
is connected.
NOTE:
Never connect the oxygen sensor to this connector
while the monitor is powered. This will damage the oxygen sensor
1. Power
Analog output
Terminal Block
Fuse
(Field
replaceable)
Common
Common
mA output
+ 24 VDC
+ 24 VDC
2. Sensor cell
connector