PureAire Monitoring Systems, Inc.
31
To change this value, push the joystick right to display the relay state. The display will
Indicate
LATCHING.
Pushing the joystick down will change the relay state from
LATCHING to NONLATCH. Press
ENTER
to accept the value.
After entering the relay state the display will default back to the
Set Latching
menu.
The display will scroll the following:
From this main menu, pushing the joystick to the right will select the sub menu and the
digital display will scroll the following:
Push the joystick twice to select the Audio Alarm relay to be adjusted. The display will
scroll;
Set Latching Audio Alarm.
This is the menu at which to adjust the Audio alarm relay state on the Air Check
O
2
Monitor.
NOTE:
The O
2
system must have the audio alarm option module installed to access
this menu. If this option is installed the display will indicate N/A, (not available)
To change this value, push the joystick right to display the relay state. The display will
Indicate
LATCHING.
Pushing the joystick down will change the relay state from
LATCHING to NONLATCH. Press
ENTER
to accept the value.
LATCHING
20.9%
NONLATCH
20.9%
..Set Latching….
20.9%
..Set Latching Audio Alarm…
20.9%
LATCH
20.9%
..Set Latching Relay 1…
20.9%