PureAire Monitoring Systems, Inc.
36
Push the joystick down to access the next sub menu;
Set Alarm 2 Hysteresis
will scroll
on the digital display. Pushing the joystick again to the right will display a value 0.0%.
Pushing the joystick up increases the percentage up to a maximum value of 2.5%.
Adjust the digital display until the desired hystereses value is selected.
Press
ENTER
to accept this value. The digital display will revert back to
Set Alarm 2
Hysteresis.
Push the joystick down to access the next sub menu;
Set Alarm Audio Hysteresis
will
scroll on the digital display. Pushing the joystick again to the right will display a value
0.0%. Pushing the joystick up increases the percentage up to a maximum value of 2.5%.
Adjust the digital display until the desired hystereses value is selected.
5.5.10 Set Sensor Adjust
This menu will permit fine-tuning of the oxygen readout to a known concentration of
Oxygen. It is recommended to adjust the oxygen display to ambient oxygen levels of
20.9%. To access this menu push the joystick down to display the
Sensor Adjustment
menu. This will scroll on the digital display.
NOTE
:
It is recommended to warm up the Air Check Oxygen monitor
for two hours before making any adjustments to the sensor.
Push the joystick right to access the first sub menu;
Set Sensor Span
will scroll on the
display. This is the menu that will permit fine adjustment of the ambient oxygen
reading to 20.9%.
Push the joystick right to access the sensor span. The display will indicate a value
between 0 and 255 counts. Pushing the joystick up increases the counts and decreases
..Sensor Adjustment…
20.9%
..Set Sensor Span..
20.9%
..Set Alarm 2 Hysteresis…
20.9%
..Set Alarm 2 Hysteresis…
20.9%
..Set Audio Alarm Hysteresis…
20.9%