PureAire Monitoring Systems, Inc.
48
How to connect a remote fan contactor to the O2 monitor
Externally powered
relay
COM
N/O
To external voltage
source, 24 VAC or
24 VAC, 2 amps max.
Relay 1, or Relay 2
Terminal Strip
External relay coil.
Relay coil voltage
rating must match
external voltage
source.
External relay using
internal 24 VDC
power supply
COM
N/O
Relay 1, or Relay 2
Terminal Strip
+24
VDC
-24
VDC
External 24 VDC
relay coil. Install
1N4004 or similar
diode across coil.
CAUTION:
PureAire Monitoring
Systems does not recommend
using 120VAC or higher voltage
inside the monitor. Use 24 VAC
or 24 VDC if possible
.