PULSAR MEASUREMENT
65
2. Install the stainless steel mounting brackets on the pipe. Position them at approximately the correct
separation distance. Exact measurement is not required at this time. Tip: Use a 5/16” nut driver to tighten
the hose clamps.
3. Use the built-in alignment bar with ruler to set the bracket spacing to the separation distance provided in
the Setup menu. One transducer is placed at the 0”/0mm reference position, and the other at the
separation distance from the Setup menu. For TMK-C2 kits, use the alignment bar extension to reach
further separation distances.