PULSAR MEASUREMENT
15
SET UP (cont.)
Vel
When Lining = Other, enter speed of sound of liner (consult
factory).
Thick
When Lining ≠ None, enter liner thickness.
Crossings
Crossings 1 = Z mounting
Crossings 2 = V mounting
Crossings 4 = W mounting
Nominal Pipe
Size, in
Recommended Crossing,
SE16A
Recommended Crossings,
SE16B
Recommended Crossings,
SE16C
0.5-1.5
4
N/A
N/A
2-3
2
2
N/A
4-6
2
2
2
8-10
N/A
2
2
12-24
N/A
2
2
26-48
N/A
1
2
Older pipes are often degraded or scaled on the inside.
These conditions can hinder the ability to receive a strong
signal when Crossings = 4 or 2. We suggest starting with
Crossings = 2 or 1, respectively, in these cases.
Zero Tare
Used to calibrate zero-flow measured by the TTFM 6.1 in
process. Flow in the pipe should be confirmed as 0 before
enabling, or significant errors in flow accuracy could occur.
Set Calibration/Damping to 0%, and under no flow
conditions and with a full pipe, select Yes to force readings
to zero.
--
-------
Setup------
r
SE16B
Senso
Angle
37
°
Fluid
Water
Temp Mode
Fixed
Temp
77.0 F
Pipe
PVC
OD
4.5000 in
Wall
0.2500 in
Lining
None
Crossings
2
Zero Tare
No
Sens Space
2.299in
Velocity
0.00 ft/s
Signal
Strength
100
%