TTFM 6.1 INSTRUCTION MANUAL
32
SE16C Pipe Preparation and Bracket Mounting
Prepare an area 2" wide by 4" long (50mm x 100mm) for each sensor bonding by removing loose paint, scale and
rust. The objective of site preparation is to eliminate any discontinuity between the sensor and the pipe wall,
which would prevent acoustical coupling. A sanding block is included with every meter to facilitate proper pipe
preparation.
A Sensor Mounting Kit is supplied with each flow meter. It includes recommended coupling tape and compound,
and a stainless-steel mounting bracket with adjustable pipe straps. Use the Alignment Bar (included) to align
sensor brackets for V and W mode mounting.
IMPORTANT:
Coupling tape is recommended for installations where the sensors cannot easily be accessed in the
future to reapply coupling compound, or in installations where the coupling compound may wear away faster
than normal (>250 deg F or where submersion is frequent). The coupling tape is subject to greater signal
attenuation than the coupling compound, so using coupling compound is recommended when it is possible to
periodically reapply the coupling in the future. If you would prefer a maintenance free installation, you can use the
coupling tape at initial installation, but may need to switch to coupling compound if you do not have a strong
enough signal.
IMPORTANT
: The SE16C transit-time transducers do not have arrows on top of them. The cable connection
points should be facing away from each other at installation.
SIDE VIEW
SIDE VIEW
SIDE VIEW
SIDE VIEW