EN
Overpressure protect
adult mode
295±5 mmHg
pediatric mode
240±5 mmHg
neonatal mode
145±5 mmHg
Resolution
Pressure
1 mmHg
Pulse Rate
±5 %
Measurement Accuracy
Cuff Pressure Accuracy
Static pressure: ±3 mmHg(±0.4 kPa)
Error
The BP Value of the device is equivalence with the measurement value
of Stethoscopy. The error meets all the conditions in the ANSI/AAMI SP-
10:2002+A1:2003 +A2:2006.
Operating temperature/
humidity
+5°C~40°C
15 %RH~85 %RH(Non-condensing)
Transport
Transport by general vehicle or according to the order contract, avoid
pounded, shake and splash by rain and snow in transportation.
Storage
Temperature: -20 ºC~+55 ºC; Relative humidity: ≤95 %; No corrosive
gas and drafty.
Atmospheric pressure
700 hPa~1060 hPa
Power Supply
DC 3 V
Battery service life
When the temperature is 23°C, limb circumference is 270 mm, the
measured blood pressure is normal, 2 "AA" alkaline batteries can be
used about 150 times.
Rated Power
≤ 3.0 VA
Dimensions
128(L)*69(W)*36 mm(H)
Unit Weight
240 gram (without batteries)
Safety classification
Internally powered equipment
Type BF defibrillation-proof applied par
Service life
The service life of the device is five years or 10000 times of BP
measurement.
Date of manufacturer
See the label
Accessories
Standard Configure:
Adult Cuff: limb circumference 25-35 cm (middle of upper arm)
Software CD, User Manual, USB data line, BP extending tube, A pouch.
Separate Sale:
Pediatric Cuff: limb circumference 10-19 cm (middle of upper arm)
Neonatal Cuff: limb circumference 6-11 cm (middle of upper arm)
8. Warranty and service
We provide 1 year warranty for material and manufacturing defects of the product.
The warranty does not apply:
- in case of damage caused by improper operation
- for wearing parts
- for defects that were already known to the customer at the time of purchase
- in case of the customer's own fault
The statutory warranties of the customer remain unaffected by the guarantee.
For the assertion of a warranty claim within the warranty period, the customer must provide proof of
purchase.
The guarantee is valid for a period of 1 year from the date of purchase against
Novidion GmbH, Fuggerstr. 30, 51149 Cologne, Germany
to assert its rights. The customer has the right to have the goods repaired by us in case of a warranty
claim.
or at workshops authorised by us.
The customer is not granted any further rights based on the warranty.