successful, and the device can be safely terminated.
Warning
If any sign of damage is detected, or the instrument displays some error messages, do not use it on any
patient. Contact biomedical engineer in the hospital or our Customer Service Center immediately.
The device can be used normally after it is turned on, without waiting for the device to be prepared.
Note
Check all the functions that will possibly be used and make sure that the device is in good status.
2.4 Connect Sensor
Note
For information on correct connection of NIBP cuff, refer to Figure 2.4.1
Figure 2.4.1 Connection Method
Connect the sensor between the Monitor and the measure part of the patient.
3. Function Interface
3.1 Main Interface
Press
to power on the instrument. The indicator will circularly flash once, which shows the boot-
strap is successful, then release the button, the system will enter the main interface.
In common user mode, if there is no button operation during the time when system sets, the device will
turn off LCD and enter into standby mode. If there is no operation in standby mode, the device will
automatically turn off after a while; the "RUN" indicator flashes once every 3 seconds to prompt the
device is in working state.
When the power is low, the battery bar is empty, at the same time the prompt sound occurs and the red
indicator flashes regularly.
In the Main Interface:
Prompt status is displayed in the left top of the screen,
button can pause the prompt status shortly.
User bar displays the current patient type (adult, pediatric, neonatal), and the amount of the common
user's data record.
Current date and time is displayed in the middle top of the screen.
Note
All interfaces display power icon, prompt switch, as well as a small font of the current time,
except the trend interface.
The oldest record will be overwritten after the memory is full. “Memory full" message is shown in
the main interface.
3.2 Measuring Interface
Measuring interface displays real-time cuff pressure and the current measurement information. In the
measurement process, all buttons are disabled, except the
and the
buttons.
Note
In any interface except the measurement interface, press
button to exit current interface and go
back to the boot-strap interface.
3.3 Measure Result Interface
The measure result includes:
SYS: systolic blood pressure (mmHg/kPa)
DIA: diastolic blood pressure (mmHg/kPa)
PR: pulse rate (bpm)
If there is an error during the measurement, an error message text will appear on the screen. If the
PROMPT SOUND is set to be on, the sound would occur. Press the SILENCE key to stop the sound and
press it once more to continue
3.4 System Menu
In the main interface, according to the text in the middle bottom of the screen, press
button, then
enter the system menu and execute different option operations by using
and
buttons.