After the ambulatory blood pressure menu is operated correctly (refer to 3.4.1), select "ABPM"
menu in [SYSTEM MENU] to enter its interface.
Switch the "ABPM ON-OFF" to "BEGIN", then a prompt message for ABPM of current user will pop
up, such as:
Figure 3.4.5 ABPM Prompt Menu
Press
button, clear the ambulatory blood pressure measure data, enter ambulatory blood
pressure mode, and start the ambulatory blood pressure monitoring. Refer to Figure 3.4.6 for the
ABPM interface.
Press
button, save ambulatory blood pressure measure data, enter ambulatory blood pressure
mode, and start the ambulatory blood pressure monitoring. Ambulatory blood pressure
measurement record includes previous data. Refer to Figure 3.4.6 for the ABPM interface.
Press
button, give up the choice, return the previous menu, and the ambulatory blood pressure
monitoring is not turned on.
5.
ABPM Working Interface
In ABPM working environment, the backlight is only active for 5 seconds. Press any button, except
, to wake the backlight. ABPM working interface is as shown:
Figure 3.4.6 ABPM Working Interface
If PROMPT SOUND occurs, press SILENCE key to stop it and press it again to continue.
In ABPM working interface, long press
button, the “exit ABPM” interface will pop up. In this
interface, press
button to exit ABPM working interface, and enter the ordinary user working
interface, the boot-strap interface will be displayed. In the “exit ABPM” interface, press
button
to exit the interface, and return to the ABPM working interface.
In the ABPM working interface, to turn off the device, exit the ABPM mode first, then long press the
power switch to turn it off.
6.
ABPM Data Review
Select "ABPM DATA" item in "ABPM" menu to enter the data review interface.
"BIG FONT" display interface: Every record is an interface, and display contents include: the
current user, total of the current user record data, serial number of the record, stored time of
the record, high pressure, low pressure, mean pressure, Pulse Rate.
In the ABPM "BIG FONT" data review interface, press
button to select "LIST", the data table
interface will pop up. Each interface contains 5 records, every record includes time, high
pressure, low pressure, mean pressure, Pulse Rate.
In the ABPM "LIST" data review interface, press
button to select "TREND", the data trend
interface will pop up. Trend interface can trace 100 trend records, if measuring data are more
than 100 items, press
,
buttons to glide trend curve for left and right, the scale of the
vertical axis and the starting point, end point automatically adjust according to the width of the
stored data. The date displayed at the bottom of trends are the data recording time of the first